56
9.8
Replacing the Printer
Caution!
Before starting, disconnect the instrument from the power source.
1. Take off the autoclave cover (see para. 9.2
“
Removing the
Autoclave’s Outer Covers
”
).
3. Disconnect the flat cable (1) connecting the printer to the DIG-T2
board.
4. Unscrew the two screws (2) fastening the printer to the panel.
5. Remove the printer and install the new one.
6. Verify that the printer is working by performing a self-test. This is
done by pressing and holding the printer feed button and turning on
the autoclave.
7. Run a cycle and verify that the printer operates correctly.