47
9
MAINTENANCE AND REPLACEMENT PROCEDURES
9.1
Safety Tests after Repair
ATTENTION!
After every repair or dismantling of the enclosure, it is
recommended that the following tests be performed:
1. Enclosure Current Leakage Test.
The test should be performed using a Megger. Make sure the
autoclave is unplugged. The electrical potential of the testing
instrument should be 500V. The insulation resistance should be at
least 2 M
Ω
.
The test is successful if there was no leakage.
2. Protective Earth Impedance Test
The test should be performed using an Ohmmeter. Make sure the
autoclave is unplugged from the wall outlet. Next measure the
resistance between the grounding screw on the rear plate (or any
other unpainted metallic part) and the grounding pin of the power
cord plug. The resistance should not exceed 0.3
Ω
, a high
resistance would indicate a faulty ground connection.
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...