57
9.9
Replacing the Door Switch
The door switch is designed to ensure that the door is properly closed
before and during a cycle.
Caution!
Before starting, disconnect the instrument from the power source
and ensure that there is no pressure in the autoclave’s chamber.
1. Take off the autoclave cover (see para. 9.2
“
Removing the
Autoclave’s Outer Covers
”
).
2. Disconnect the wires from the door microswitch.
3. Remove the microswitch and replace it with a new one.
4. Reconnect the black ground wire (2) to the microswitch.
5. Test the connection with an ohmmeter. Connect the ohmmeter to
the common terminal of the microswitch (1) and chassis ground. In
the “door open” position the ohmmeter should show no continuity
and in “door close” position the ohmmeter should show a complete
circuit.
5.1 If it fails the test then check that the ground wire is connected
to the correct terminal. If it is then replace the microswitch.
5.2 If it passes the test then connect the green wire from the
electronic box to the common terminal (1) of the switch
6. Reassemble the autoclave cover.
3
2
1
4
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...