113
Corrections
1.1
Disconnect the pipe from the pump, let
water flow to remove the air pocket, and
then
reconnect the pipe.
1.2 Clean
pipe.
1.3
Clean the filter
See section 10.4.26
See section 10.4.27
4.1
Clean the filter screen. Use caution the
filter cap may be HOT!
4.2
Check out the valve according to section
10.4.4 and 10.4.36
4.3
Check the line and clear any obstruction.
4.4
Replace the no-return valve.
Possible cause check-up and tests
Check pipes connecting the water reservoir, filter
and water pump.
1.1
There is an air pocket at the pump inlet.
1.2
Pipe is clogged.
1.3
Filter is clogged
See section 10.4.26
See section 10.4.27
4.1
The filter screen located on the side of
the autoclave may be clogged.
4.2
The fast exhaust valve (73) may be
clogged or defective.
4.3
The exhaust line may be clogged
4.4
The no-return valve may be defective.
10. TROUBLESHOOTING (cont.)
10.6
Water pump malfunction
Symptom
1.
Water pump is operating
but no water is flowing or
the rate is too slow.
2.
Pump does not work.
3.
The pump works
continuously.
4.
Water does not exit the
chamber.
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...