54
9.6
Replacing the Pressure Gauge
Caution!
Before starting, disconnect the instrument from the power source
and ensure that there is no pressure in the autoclave’s chamber
and jacket.
9.6.1
Model 2540
1.
Take off the door cover (see para. 9.18 Replacement of
the Door Cover).
2.
Remove the pressure gauge from the door by unscrewing
it from the door
3.
Install the new pressure gauge using Teflon tape to seal
the threads. Verify that the gauge’s tube does not
protrude from the doors inner surface.
4.
Operate the autoclave and verify that there are no leaks.
5.
Reinstall the door cover.
9.6.2
Model 3870
The gauge is located in the plastic housing adjacent to the door
and directly in front of the electronic box.
1.
Take off the autoclave cover (see para. 9.2 “Removing
the Autoclave’s Outer Covers”).
2.
Disconnect the copper tube at the back of the gauge.
3.
Remove the pressure gauge.
4.
Install the new pressure gauge.
5. Reconnect the copper tube. (The copper tube uses a
compression fitting so no
Teflon tape is needed).
6.
Operate the autoclave and verify that there are no leaks
7.
Reassemble the autoclave’s cover
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...