103
Corrections
29.1
Replace the DIG-T2 board.
29.2
Fix the connections.
29.3
Disassemble the valve and check that the
plunger is free to move and that the seat
is clean and free from debris.
29.4
Replace the ANL-T2 board.
Possible cause check-up and tests
The problem may be a mechanical failure of the
valve, the valve driver located on the ANL-T2
board or the control circuit on the
DIG-T2 board.
29.1
Check TP12. If the voltage is higher than
4V and the valve is open, then the DIG-
T2 board is faulty.
If the voltage at TP7 is <1VDC and the
valve is open, the problem is not the
DIG-T2, but some other element of the
system.
29.2
Check the connection between the
connector JP10 – 8 and the valve for a
ground short.
29.3
Remove the cover protecting the
terminals of the coil and check the
voltage across the coil when the solenoid
is connected. If the coil voltage is less
than 2V and the valve is open, there is a
mechanical problem.
29.4
If the voltage across the coil is higher
than 10VDC, then the ANL-T2 board is
faulty.
10. TROUBLESHOOTING
(cont.)
10.4
In Process Malfunction
(cont.)
Symptom
29.
Steam valve (93) is always
open.
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...