53
9.5.2
Replacing the drain valve assembly
1.
Take off the autoclave cover (see para. 9.2
“
Removing
the Outer Covers of the Autoclave
”
).
2.
Disconnect the drain tube from the valve, using a 9/16
”
wrench.
3.
Remove the nut (3) and the “ring for drain valve” (2).
4.
Remove the drain valve (1) from the panel.
5.
Install a new valve according to the drawing below.
6.
Verify that there is no leakage.