24
— Press
the
PROGRAM
key to advance to the next
value.
— The
cursor
on
the
bottom
row
will
move one
position to “R”;
now
enter the pressure value
recorded from the digital display to which you
have already added 14.5.
— Press
the
PROGRAM
key one final time
If PROGRAM WAIT has not been displayed then
continue to press the
PROGRAM key
until it appears.
After a brief period the unit will return to a Ready
state
This completes the calibration procedure for
temperature e and pressure. A sterilization cycle
should be performed to verify that the system is
calibrated and running properly.