42
8.2.15
- Overrides the pressure calculated for sterilization
This parameter is used to insure that during sterilization the
temperature is correctly maintained. It does this by raising the
value of the sterilization pressure necessary to maintain a
sterilization temperature. The sterilization pressure is calculated
by the unit based upon the
SteTemp
parameter.
For example; if the required temperature is 250ºF, the required
pressure calculated by the program is 16 psig.
This pressure can be increased by changing the value of
. If this value is set to ø, the system will be maintained
at 16 psig during the sterilization stage if this value is set to 1.5,
the system will be maintained at 17.5 psig in the sterilization
stage.
Access
Code
-
13
Resolution -
0.1
psi
Minimum
Value
-
Ø
Maximum Value -
3 psi
Cycle
1 2 3 4 5 6
Default
Values
Value
0.8 0.8 0.6 0.8 0.8 0
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...