38
8.2.5
Ex.Mode – Rate of steam discharge during the exhaust stage
This parameter sets the rate at which steam is discharged during
the exhaust stage.
Access
Code
-
13
Resolution -
1
Value
-
1 or 4
Fast Exhaust
= 1
Slow Exhaust
= 4
Cycle
1 2 3 4 5 6
Default
Values
Value
1 1 1 1 1 1
8.2.6
End Temp – Cycle End Temperature
This parameter sets the finish temperature of the exhaust stage.
The unit will signal a Cycle End only when the exhaust timer
has reached zero and the unit is below the
End Temp.
This also
indicates the proper time to open the door and unload the
chamber.
Access
Code
-
13
Resolution -
1ºF
Minimum
Value
-
104ºF
Maximum
value
-
280ºF
Cycle
1 2 3 4 5 6
Default
Values
Value
273 273 250 273 273 280
8.2.7
Puls num
.
–
No. of vacuum pulses in the prevacuum stage
This parameter sets the number of vacuum pulses during the
prevacuum stage.
Access
Code
-
13
Resolution -
1
Minimum
Value
-
1
Maximum
Value
-
7
Cycle
1 2 3 4 5 6
Default
Value
Value
2 4 4 4 4 1
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...