40
8.2.11
VacTime2
–
Waiting time at the end of each of the remaining
pulses
This parameter sets the wait time at the end of each of the
remaining vacuum pulses, during the prevacuum stage. After
reaching
VacDip2
, there is a waiting time to allow the unit to
stabilize. This time is defined as
VacTime2
and during this time
the vacuum system continues working.
Access
Code
-
13
Resolution
-
1
second
Minimum Value
-
3 seconds
Maximum Value
-
1800 seconds (30 minutes)
Cycle
1 2 3 4 5 6
Default
Values
Value
3 3 3 3 3 3
8.2.12
WtrResTime – Timed extension for pumping mineral free
water
This parameter sets the length of time to continue pumping
mineral free water into the steam generator after the upper
electrode senses water.
This is necessary to prevent a situation where the water pump
would cycle on and off as the water level rises and falls around
the upper water-sensing electrode.
Access
Code
-
13
Resolution
-
1
minute
Minimum Value
-
Ø minutes
Maximum Value
-
90 minutes
Cycle
1 2 3 4 5 6
Default
Values
Value
5 5 5 5 5 5
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...