Corrections
—
Repair or replace the
faulty item.
1.5
—
Tighten loose screws.
—
Reconnect
grounding
cable.
Possible cause check-up and tests
—
Turn the power back on and check for 5V
and 12V on the Test Point board, if
voltages are okay, then one connector at a
time reconnect the connector and test the
voltage. Each time turning the power off
before reconnecting.
—
If a cable or connector causes a problem,
but the cable or connector are, good then
proceed down the line to the device
attached to the other end of the cable or
connector and continue with the previous
step. The problem could
be with any one of the external devices
connected to the system (valves, sensors,
etc.) or another electronic board.
1.5
Check the grounding
—
Connect the negative probe of the
multimeter to TP1 on the Test Point board.
—
Using the positive probe check the voltage
at various grounding points on the
autoclave, on the ANL-T2 and DIG-
T2 boards, on the enclosure of the electric
box and the main grounding screw of the
machine.
10. TROUBLESHOOTING (cont.)
10.3 Pre-process malfunction
(cont.)
Symptom
1. (cont.)
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...