70
9.19
Replacing the Closing Device
Caution:
Before starting, verify that there is no pressure in the autoclave
chamber.
1. Remove the “C” clip (9) from the bottom of the door locking pin
(6). Newer models will have a cotter pin.
2. Remove the locking pin (6) by pulling it up out of its position.
3. Remove the closing device assembly. Be careful not to lose the two
Teflon disks (10) resting on the top and bottom of the lock housing
axe (5).
4. Position the new closing device on the lock housing axe (5),
making sure that the Teflon disks (10) are in between the lock
housing axe (5) and the screw housing (3).
5. Insert the pin (6).
6. Replace the “C” clip or cotter pin (9).
CLOSING DEVICE
No.
Description
No.
Description
1
Bushing
6
Door locking device pin
2
Door tightening bolt assembly
7
Bakelite handle
3
Locking screw housing
8
Closing bridge “c” clip
4
Locking base
9
Cotter pin
5
Locking housing axe
10 Teflon disk
Summary of Contents for 2540 EHS
Page 10: ...9 REAR VIEW...
Page 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Page 19: ...18 AC T1 BOARD Top View Bottom View...
Page 30: ...29 6 4 Input Output Graph...
Page 31: ...30...
Page 32: ...31...
Page 33: ...32...
Page 34: ...33...
Page 35: ...34...
Page 66: ...65...
Page 68: ...67...
Page 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Page 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Page 126: ...125 PIPING DRAWING FOR 2540 EHS...
Page 127: ...126 PIPING DRAWING FOR 3870 EHS...
Page 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Page 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Page 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...