Page 60 of 117 pages
The next screen will appear as such
Where:
T1 – T4 display the temperature received from T1 until T4 (without a filter).
D1 – Displays the pressure in gasket 1
Gen – Displays the pressure in the generator.
In this stage you have to calibrate the displayed sensors.
Press on ‘UP’ to move to the next stage, or wait 15 seconds and it will do so
automatically.
The next screen to appear refers to the digital inputs.
As with the digital outputs, it will progress to the next value automatically after 15
seconds, or by pressing the UP key. Values 5 & 7 – 11 will be OFF.
The final screen to appear in the INOUT test gives 2 options:
Pressing UP will return to the Code Entry screen for the INOUT Test.
Pressing STOP will return the display to the ST.BY screen.