4.3.4 Changing Cycle Parameters
The screen that allows for parametric changes appears as follows:
Pressing Clear or Enter before entering the code returns to the Stand By screen,
and the RUN & FAIL LED's are lit.
Enter the correct code only via the UP/DN keypads, press ENTER to proceed to
the next screen.
Pressing the UP/DN keypads can also change all parameter values.
Cycle Setup Parameters:
Code 1 or 11 Parameters
STE Temp – Required sterilization temperature for the process
Entry code
1 or 11
Resolution
34ºF
Minimum value
221ºF
Maximum value
280ºF
STE Time – Required sterilization time for the process
Entry code
1 or 11
Resolution
1 minute
Minimum value
Ø minutes
Maximum value
59 minutes
Dry Time – Required drying time for the process:
Entry code
1 or 11
Resolution
1 minute
Minimum value
Ø minutes
Maximum value
59 minutes
Heat T.O. – Maximum time for heating the chamber for starting sterilization
When the duration of the heating stage is longer than Heat T.O., the program will
collapse with the message ‘Low Heat’.
Entry code
11
Resolution
10 seconds
Minimum value
1200 seconds (20 minutes)
Maximum value
3600 seconds (minutes)
Summary of Contents for 2301006
Page 2: ......
Page 14: ...Page 9 of 117 pages...
Page 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Page 79: ...Page 74 of 117 pages Control Panel Assembly...
Page 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Page 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Page 106: ...Page 101 of 117 pages...
Page 107: ...Page 102 of 117 pages...
Page 109: ...Page 104 of 117 pages XPCS Manual...