Program: 9- Slow 250
Front door close Msw
PT100 Drain Control 5
PT100 Condense 1
PT100 Condense 2
IMT Jacket
IMT Chamber
Open front door
Close front door
Ejector
Buzzer
13 – Cool drain
15 – Water to ejector
43 – Air inlet
73 – Fast exh
74 – Slow exh.
75 – Vacuum
97 – Steam to jacket
93 – Steam to chamber
273ºF / 310kPa
250ºF / ~ 210kPa
199 kPa
230ºF / ~ 115kPa
100kPa
50kPa
25kPa
0kPa
Start (Close door)
Door secure
Temp. control
Pres. control
On (activated)
115kPa
Ex. Mode 4
250ºF 15min.
Condition to open
T<230ºF ; P<115kPa