Page 68 of 117 pages
6.
TRANSPORTATION OF THE AUTOCLAVE
To facilitate transporting the autoclave through narrow passageways, the
cabinet is made in two sections. After transporting the autoclave to its
destination reassemble.
1.
Remove all the bolts which tighten the control panel with the
autoclave panel (2,3).
2.
Remove the bolts that tighten the control base to the autoclave base
(1).
3.
Disconnect the multi-pins connector mounted on the main electric
box.
4.
Separate the two autoclave parts.
After the autoclave has been installed at the desired location, connect the
two parts according to the reverse order of the directions given for
dismantling.