F. O. Mode
This mode enables F.O. calculations, the processing being based on data fed by
two temperature sensors one mounted in the chamber, one on the condensate
pipeline.
The F.O. calculation is performed for cycles with Ster. Temp. in the range of
215ºF - 273ºF, especially used in Programs 9-12.
F.O. = 0 (unable),
F.O. 1= total heating time + total cycle time + cooling time.
F.O. 2 = compensates the difference between heating time,cooling time and
exhaust. Calculates using factor 1.5 of the total accumulated heating and cooling
time and converts that into the equivalent sterilization time.
Temp. F.O. = Determines the temp. at which F.O. calculations should begin.