Page 113 of 117 pages
•
When download process is completed, the message "Download
Finished" will be displayed in the status bar.
Note: If error occurs during download process, relevant error massage will
be displayed. The message box with 'Retry' option will be displayed.
After completion of the downloading process, the download window will be
displayed as follows:
Note: Upload and Download green messages of the Parameters and Gain &
Offset tasks are displayed only if the relevant options were selected in the
'Options' window
.
Expectable errors list
Can not upload Gain and Offset!
Description:
This message may be displayed during download process, when
the program is trying to upload Gain and Offset values.
Reason:
Probably a communication failure between PCS Target and the PC.
Tips:
Check if communication RS232 cable is connected. If it is a modem
connection, check if Modem cable is connected in this case it can be that the
remote modem cable is disconnected or RS232 cable between the remote
modem and target platform is disconnected.
Summary of Contents for 2301006
Page 2: ......
Page 14: ...Page 9 of 117 pages...
Page 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Page 79: ...Page 74 of 117 pages Control Panel Assembly...
Page 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Page 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Page 106: ...Page 101 of 117 pages...
Page 107: ...Page 102 of 117 pages...
Page 109: ...Page 104 of 117 pages XPCS Manual...