Page 52 of 117 pages
Doors Number
After entering the language, the display is as follows:
The options for this autoclave having one door, no key has to be selected. Press
ENTER, and the system moves on to the next screen.
Door Type
There are 3 different door types for autoclaves – Manual Door, Hinged Automatic
Door, or Sliding Automatic Door.
In this case the door is automatic, just press ENTER as this is the option already
displayed (ØØØ1). Enter these values either through the UP/DN keys and then
press ENTER, or physically enter the digits whereby the screen will automatically
progress to the next display after entering the fourth digit.
Atmospheric Pressure
The atmospheric pressure at Sea Level is 100kPa. For every 100m above sea
level, the atmospheric pressure drops 1kPa, and for every 100m below sea level,
the atmospheric pressure increases 1kPa.
If pressure is set correctly, press ENTER and the next screen appears. If pressure
needs to be altered, this may be done either through the digit keypad or the
UP/DN keypads.
Jacket (Jacket Type)
If there is no Jacket, then press Enter.
If there is a Jacket, then via the UP/Dn keypads, choose the correct number
according to the following conditions:
♦
1-There is a Jacket – the steam to the chamber passes from the jacket
♦
2-There is a Jacket – the steam to the chamber passes directly and not via
the Jacket. Press 2 for this autoclave.
♦
3-There is a Jacket and it is divided without a separation from the steam
generator, and the steam passes via the jacket to the chamber.
Summary of Contents for 2301006
Page 2: ......
Page 14: ...Page 9 of 117 pages...
Page 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Page 79: ...Page 74 of 117 pages Control Panel Assembly...
Page 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Page 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Page 106: ...Page 101 of 117 pages...
Page 107: ...Page 102 of 117 pages...
Page 109: ...Page 104 of 117 pages XPCS Manual...