INSTALLATION INSTRUCTIONS
Only qualified and authorized personnel are allowed to install, check,
test, maintain and serve the autoclave.
Mounting
♦
Check after unpacking the autoclave that the Serial Number (S/N) on its
nameplate corresponds with the package number.
♦
Place the autoclave on a level surface, leaving adequate space around it
for operation and service requirements. (see also the drawings)
♦
The ambient temperature around the autoclave should not exceed 104ºF,
80% humidity and the room must be ventilated 10 air changes/hour.
♦
Do not use the autoclave in presence of dangerous gases and vapors.
Utilities
Connect the autoclave to the utility supplies as follows:
Air
♦
Connect the air connector on the autoclave to the building air source by
means of a 1/2” pipe at 90 to 120 psi (6 to 8bar)
♦
If there is no external air source, use a small, independent air
compressor.
Mineral Free Water
♦
Connect the water reservoir container to a free mineral water supply
source by a 1/4".
See also the Water Quality chapter!
Tap Water
♦
The autoclave should be supplied with tap water; its hardness should not
exceed 0.7-2mmol/l. See also 2.5 Water Quality.
♦
The tap water pressure should be between 2 to 4 bar (30 to 60 psi).
♦
Connect the feed water to the network by a 1/2” pipe.
Summary of Contents for 2301006
Page 2: ......
Page 14: ...Page 9 of 117 pages...
Page 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Page 79: ...Page 74 of 117 pages Control Panel Assembly...
Page 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Page 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Page 106: ...Page 101 of 117 pages...
Page 107: ...Page 102 of 117 pages...
Page 109: ...Page 104 of 117 pages XPCS Manual...