Page 67 of 117 pages
We are interested in correcting the measured temperature for 273
º
F without
changing the temperature at 250
º
F which is accurate.
Considering the above example, setting data can be done as follows:
Upper Row
A 140º F
R 141ºF
A 266º F
R 266ºF
The system calibrates the new offset and gain and loads them into the
non-erasable memory.
As a result, while running a 273
º
F program the correction is automatically
done, while at 250
º
F no change is felt.