Page 58 of 117 pages
F.O. Temp Mode
Determines the temperature F.O. calculations will begin.
Entry code
13
Resolution
1
Minimum value
215
Maximum value
273
Select PT100
There are 3 sensors that regulate cycles – one for condense, one in the chamber,
and one in drain control output.
Selections:
1. Sensor on the condense line x 2 (dual). (2)
2. Sensor in the chamber (1)
3. Sensor for drain control (5).
If sensor number 2 is not connected (faulty), the system will automatically work
with sensor no.1 (the sensor in the condenser), even if 2 was selected, as 2 is only
an option.
This is the final parameter under Code 13. Pressing ENTER; the display returns
to the ST.BY screen.