Page 59 of 117 pages
4.3.5 Test Programs
To arrive at this option, return to the ST.BY screen and press the OPTION Keypad
(the middle command keypad). Then screen will appear as with SETUP and
Calibration. The word above the right command keypad will be INOUT. Press the
right keypad you will be asked to enter a code.
Via the numeric keypad, enter the correct code. The display will automatically
move on to the next screen after entering the final digit.
The screen will appear as follows:
Every 15 seconds the Digital Out number will automatically increase by one until it
reaches Digital Out21. It then automatically passes on to the next screen.
Pressing the UP keypad will also progress through the Digital Out values.
The next screen to appear is the following:
The screen will display the message ‘Electrd – h # #’. # # is a number value.
When the electrode touches water the value will be small, from 80 down to 0 when
the water is high in electric molecules. When the electrode is in the air (not
touching water) the value will be higher than 80 up to 99.
This electrode is located in the generator so that if there is no water in the
generator the value is said to be 99 (greater than 80). This electrode is the
highest electrode in the generator.
Press on ‘UP’ to move to the next stage, or wait 15 seconds and it will do so
automatically.
The next screen to appear will display the message ‘Electrd – L ##’, with ## being
a number value. When the electrode touches water the value will be small, from 80
down to 0 when the water is high in electric molecules. When the electrode is in
the air (not touching water) the value will be higher than 80 up to 99.
This electrode is located in the generator so that if there is water in the generator
the value is said to be 0 (lower than 80). This electrode is the lowest electrode in
the generator.
Press on ‘UP’ to move to the next stage, or wait 15 seconds and it will do so
automatically.
Summary of Contents for 2301006
Page 2: ......
Page 14: ...Page 9 of 117 pages...
Page 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Page 79: ...Page 74 of 117 pages Control Panel Assembly...
Page 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Page 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Page 106: ...Page 101 of 117 pages...
Page 107: ...Page 102 of 117 pages...
Page 109: ...Page 104 of 117 pages XPCS Manual...