Creating a Method
Creating a MS Instrument Method
Thermo Scientific
TSQ 8000 Evo User Guide
69
Figure 68.
Acquisition-Timed Scan Settings
i.
The
Window Optimization
pane allows access to the window optimizer
settings. When the optimize button is checked, acquisition windows will be set
automatically based on the acquisition window and dwell time targets set in this
pane. For complex SRM or SIM methods, such as a method containing more
than three hundred SRM transitions, this option will help ensure a method is
created that can achieve the requested scans per peak.
Algorithm Details: If the
Optimize
checkbox is checked, the SRM acquisition
windows in the method are set to the
Desired Window
unless the
Desired Min
Dwell Time
cannot be met with the number of
Requested Scans Per Peak
. If
this occurs, then the acquisition windows are reduced until either the
Desired
Min Dwell Time
is met, or the
Minimum Window
is reached. If the
Minimum
Window
is reached first, the
Minimum Dwell Time
is reduced until the
Requested Scans Per Peak
is achievable. If the absolute minimum dwell time on
the instrument, which is 0.5 ms, is reached before the requested points across the
peak are achieved, the
Minimum Window
is lowered until the
Desired Scans
Per Peak
criteria are met or until the absolute allowed minimum window on the
instrument is reached, which is 0.24 min. In this very rare case you must reduce
the number of
Desired Scans Per Peak
, increase the
Min Baseline Peak Width
,
or reduce the number of transitions contained in your method before you will be
allowed to save your method (see
Scan Settings