5
Optimizing Your Method
Modifying an Automatic Tune
104
TSQ 8000 Evo User Guide
Thermo Scientific
Figure 103.
Changing a Tune File Name
c.
Click
OK
.
To modify advanced tune settings
1. To configure the
Q1 Lenses
, you must select the
Show Advanced Settings
check box.
The lens tune is the main portion of the tune algorithm. In this section you can choose
which components to tune, which mass to optimize for the tune, what the range of
allowed values is, how to move through the range, how much the signal must change for a
new value to be selected, and what (if anything) can be done about the resolution of the
peak and any errors that occur. Select the
Tune Lenses
check box to access the data table.
See
.
Figure 104.
Tuning the Q1 Lenses
a.
Device
—Click in this box to select the component you want to tune according to the
settings in the other columns.
Note
The following instructions refer to advanced tuning settings that most users will
never need to access. Contact your local Thermo Fisher Scientific service organization
before adjusting these settings.