Thermo Scientific
TSQ 8000 Evo User Guide
31
3
Tuning
Tuning will improve the performance of your TSQ 8000 Evo system. For optimum stability,
you must wait until the power, vacuum, and heaters lights on the front of the instrument are a
solid green. These lights indicate that the instrument has reached vacuum and that it is at the
last set temperature. If the system has been powered off for a period of time (that is, a cold
system), the system components take up to 4 hours to thermally stabilize after reaching the
temperature setpoint. If you did not vent the instrument (that is, the system is hot), the
components take approximately 10 minutes to thermally stabilize after reaching the
temperature setpoint. If you exchanged the ion source without venting, the source
components take approximately 30 minutes to stabilize temperature.
You can tune the TSQ 8000 Evo system by using the TSQ Series Dashboard.
Accessing Auto Tune
To access Auto Tune choose one of these two methods
• Double-click the
TSQ Series Dashboard
shortcut on your desktop to open the
TSQ Series Dashboard and then click
Auto Tune
. See
.
Contents
•
•
•
Updating Tunes for New RF Lens
IMPORTANT
Be sure to give the TSQ 8000 Evo system enough time to stabilize.
Otherwise, you may see mass drift, mass spectral changes, or changes in the fragmentation
of your data.
Note
If you are running samples with heavy matrix, running the samples before the
source has had time to stabilize at high temperature will prematurely dirty the ion volume
and optics.