Creating a Method
Creating a MS Instrument Method
72
TSQ 8000 Evo User Guide
Thermo Scientific
v.
Under
Acquisition Options
, select the
Allow for Asymmetric Acquisitions
check box to add extra time to the beginning or end of an acquisition window
without affecting other timing in your scan. When you select this option,
Pre-width and Post-width
columns are added to your method. Enter the extra
times in these columns.
Select the
Allow Dwell Time Prioritization
check box to increase the dwell
times for selected scans. The choices for each scan are
Normal
or
High
. Giving a
scan high priority increases its dwell time by the value you set in the
High
Priority Multiplier
box.
vi. Under
Full Scan
, you indicate if a Full Scan is to be run along with SIM and
SRM. The Mass Range, Scan Time, Start and End Time can be set after the
Use
Full Scan
box is checked. The full scan time will reduce the SRM/SIM time
without increasing the total scan time. If you only want to use full scan for part
of your method, you can enter full-scan start and end times.
b. Use the options in the
Instrument Settings
panel to further adjust your method.
i.
Use
Tune File
to select a tune file or files to be used for this scan. If you are using
EI, only the
Tune File(+)
pull-down menu is available. If you are using negative
CI, select a tune file from the
Tune File(-)
pull-down menu. Choose
AutoTune_NCI
to use the most recent negative CI automatic tune
file you
created in
If you are using positive CI, select a tune file
from the
Tune File(+)
pull-down menu. Choose
AutoTune_PCI
to use the most
recent positive CI automatic tune file you created in
software defaults to the most recent
AutoTune_NCI
or
AutoTune_PCI
tune file
you created.
ii. In the
Detector Gain
area, set the detector gain. Select the
U
se Last Tuned
Detector Gain
option to indicate that you want to use the detector gain set in
Auto Tune. If you do not need to use the gain set in Auto Tune, then you can set
the gain manually. Higher gains give larger signals, but may shorten the lifetime
of the detector when concentrated samples are detected. To manually set the
detector gain, select the
Use Specified Detector Gain
radio button and enter the
desired value in the Detector gain box.
iii. Use the
Emission Current
box to set the emission current used during the
acquisition. For optimal analytical performance and stability, use the emission
current at which the system was tuned. However, if you want to use a different
emission current, select the
Use Specified Emission Current
radio button and
enter a value in the
Emission Current (μA
) box. A high emission current will
lead to the production of more ions, but the interaction of too many ions in the
source can cause a degradation in the resolution and signal. The margin of error
is ± 0.5 μA.
Note
This option is only available when the method optimizer is not active.
Tip
Click the arrow at the top right of the
Scan Settings
panel to collapse or
expand the panel.