Tuning
Updating Tunes for New RF Lens
Thermo Scientific
TSQ 8000 Evo User Guide
43
9. Choose
3
from the
Spectra
drop-down menu. See
.
Figure 43.
Selecting the Correct Number of Spectra to Scan
10. Select
EI
from the
Cal. Gas Level
menu. See
.
Figure 44.
Setting the Calibration Gas Level to EI
11. Click
Start Scan
. Check the intensities for the masses 69, 219, and 502. Set the masses
using the
Mass
drop-down menu below each spectrum. See
.