5
Optimizing Your Method
Modifying an Automatic Tune
Thermo Scientific
TSQ 8000 Evo User Guide
103
b.
Acquisition Threshold
—Enter a minimum peak height for the full scan spectrum
on the tune report. If your peak has an intensity that is below this threshold, it will
not be stored.
c.
SIM Masses
—Select masses to be displayed on your tune report. You can select a
maximum of five masses, one from each list. At least one mass must be displayed on
the report.
9. Once you are finished configuring all the tabs, click
Save
to save the tune type. Your new
tune type appears in the Tune Types window.
10. Click
Close
to return to the dashboard. You can select the tune type in the automatic
tune window.
11. If you want to rename a tune you created, open the Tune Types dialog box by clicking
Tune Types
on the TSQ Series Dashboard.
a. Select
Rename
in the Tune Types dialog box.
Figure 102.
Renaming a Tune Type
b. The
Rename Tune Type
dialog box opens. Assign a valid new name to your tune.
You will not be able enter a name that is an invalid file name or already in use. See
.
Tip
Factory tunes cannot be altered. If a newly created tune does not perform as
expected, you can always go back to a factory tune, copy it, and change the
settings as described above. Save the new tune with a new name before closing the
Edit Tune Type dialog box.
Note
You cannot rename a factory tune.