Changing the Column
Connecting the Column to the Transfer Line
28
TSQ 8000 Evo User Guide
Thermo Scientific
Figure 27.
Positioning the Septum
14. Pull the column back from the transfer line nut. Do not move the septum from its
position on the column.
Figure 28.
Pulling the Column Back from the Transfer Line Nut
15. Loosen the transfer line nut from the column measuring tool.
16. Remove the column, transfer line nut and ferrule from the column measuring tool,
making sure not to move the septum from its location on the column.
17. Insert the column into the transfer line.
Transfer Line
Nut
Septum
Column
Measuring
Tool
Transfer Line
Nut
Septum
Column
Measuring Tool
Column