1
Introduction
Configuring Your Instrument
Thermo Scientific
TSQ 8000 Evo User Guide
11
changed the IP address using Instrument Configuration. For security purposes, you may
also want to modify the TCP port. Consult your local IT Department for help.
Figure 10.
Setting up the Network
12. Click the
Maintenance Intervals
tab to set the number of days until you plan to perform
maintenance on certain components of your GC/MS system. See
Figure 11.
Setting the Maintenance Intervals
13. Select the
Foreline Pump Oil
check box to enable the maintenance intervals. Then set
the number of days after which you want to be reminded to check the oil. The
manufacturer recommends changing the oil every 125 days. Select the
Turbo Pump
Oil/Bearing
check box to enable the maintenance reminder. Then set the number of days
after which you want to be reminded to check the oil. Refer to the Turbo Pump manual
for the manufacturer's recommended maintenance intervals.
14. Select the
Filament 1
check box to enable the maintenance reminder. In a leak-free
system, expect the filament to last between 30-360 days, depending on usage.