3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
318
ISQ 7000 Hardware Manual
Thermo Scientific
PC Communication Board, Controller Interface Board, and Support Bracket
To replace the PC communication board, controller interface board, and support
bracket
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. Use a T20 Torxhead screwdriver to remove the two screws holding the stainless steel
support bracket to the middle of the instrument.
4. Replace the bracket if necessary.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Note
If you only need to replace the support bracket, reattach it to the instrument, and
restart your system by following the instructions in
.