
3
Advanced Maintenance
Replacing Components of the Manifold
246
ISQ 7000 Hardware Manual
Thermo Scientific
8. Set the valve body and all of its components aside.
9. Remove the source magnets from the back of the manifold door and set them aside.
10. Twist and remove the front door alignment pins that hold the magnet yoke to the back of
the manifold door.
Figure 256.
Removing the Front Door Alignment Pins
Figure 255.
Removing the Source Magnets
Source Magnets
Front Door
Alignment Pins
Magnet
Yoke