3
Advanced Maintenance
Replacing Components of the Ion Source
Thermo Scientific
ISQ 7000 Hardware Manual
213
4. Use a T20 Torxhead screwdriver to remove the two screws on the back of the ion source.
5. Slide the source interface board and ion source out from under the tray without damaging
or disconnecting the wires.
Figure 201.
Disconnecting the Ion Source from the Analyzer Tray
Figure 202.
Removing the Ion Source from the Analyzer Tray
Screws
Wires