3
Advanced Maintenance
Replacing Components of the Ion Source
Thermo Scientific
ISQ 7000 Hardware Manual
191
15. Reattach the thumbscrew-repeller spacers.
16. Attach the source-repeller thumbscrews to the endcaps and twist them back on. When
you are tightening the thumbscrews, make sure the spacers stay in the groove or you could
crush them.
Figure 172.
Reattaching the Thumbscrew-Repeller Spacers
Figure 173.
Reattaching the Source-Repeller Thumbscrews
Thumbscrew-Repeller Spacers
Source-Repeller Thumbscrews