Advanced Maintenance
Replacing Components of the Ion Source
Thermo Scientific
ISQ 7000 Hardware Manual
197
4. Use forceps to remove and replace the thumbscrew-repeller spacers.
5. Reattach the source-repeller thumbscrews to the spacers.
6. Reattach the front manifold door and four screws.
7. Restart your system by following the instructions in
Figure 181.
Replacing the Thumbscrew-Repeller Spacers
Note
When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crush the spacers.
Thumbscrew-Repeller
Spacers