3.
In the Assays step, select the assay or assays that you want to use in the run.
Note:
To create a new assay, see Chapter 3, “Create and manage assays
IMPORTANT!
Ensure that you select the assay that corresponds with the sample
type that you will use in the run. If you select a wrong assay when you create a
run plan, the instrument will use incorrect settings during the run, resulting in
invalid sequencing results. To determine which system-installed assay to use for
your sample type, see “System-installed assays“ on page 30.
If you selected the Upload BAM files to Ion Reporter
™
Software reporting
option in substep 2b, make the following selections from the dropdown list in the
Ion Reporter
™
Workflow column in the row of each assay that you selected.
• Select Upload Only to upload sample data to Ion Reporter
™
Software
automatically upon run completion.
• Select the desired Ion Reporter
™
analysis workflow to upload sample data and
launch an analysis in Ion Reporter
™
Software automatically upon run
completion.
Note:
In order for the Ion Reporter
™
analysis workflow to appear in the list,
you must tag the analysis workflow for use with the IonReporterUploader
plugin. For more information, see “Tag an Ion Reporter
workflow for use with the IonReporterUploader plugin“ on page 140.
Chapter 5
Create and manage run plans
Create a run plan for a library run
5
Genexus
™
Integrated Sequencer User Guide
65