Cleaning is normally performed automatically at the completion of the previous
sequencing run. Perform a Clean instrument procedure if the sequencing run:
• was aborted or had a power failure during a run, or
• the normal post-sequencing run cleaning was not completed.
1.
In the Settings menu, tap Clean instrument.
The cleaning procedure starts automatically.
2.
Follow the on-screen instructions, then tap Next. For more information, see
“Clear the instrument deck and perform a UV Clean“ on page 81.
The user interface returns to the Home screen when the cleaning is complete.
The System Tools menu enables you to upload instrument diagnostics, manage data,
and shut down or reboot the instrument.
Perform a
Clean
instrument
operation
System Tools
Appendix B
Touchscreen reference
Settings
B
124