10
EVOS
™
M5000 Imaging System User Guide
2.
Installation
Operating environment and site requirements
•
The dimensions of the EVOS
™
M5000 Imaging System are 18 × 23 × 18 in
(W×H×D) (46 × 59 × 46 cm). The system requires a benchtop of approximately
36 × 36 in (92 ×92 cm).
•
If the system includes the optional EVOS
™
Onstage Incubator (Cat. No.
AMC1000), then add 16 in (40 cm) to the width of the bench.
•
Allow at least 5 cm (2 in) free space at the back of the instrument to allow for
proper ventilation and prevent overheating of electronic components.
•
Place the EVOS
™
M5000 Imaging System on a level surface away from
vibrations from other pieces of equipment. Tabletop centrifuges, vortex mixers,
and other laboratory equipment can vibrate the instrument during a run and
cause a decrease in instrument performance.
•
The EVOS
™
M5000 Imaging System should be installed away from direct light
sources such as windows. Ambient light can enter the imaging path and affect
the image quality.
•
Operating temperature range: 4°–32°C (40°–90°F).
•
Relative humidity range: 0–90%.
•
Operating power: 100–240 VAC, 1.8 A
•
Frequency: 50–60 Hz
•
Electrical input: 12 VDC, 5 A
IMPORTANT!
Do not position the instrument so that it is difficult to turn off the
main power switch located on the back of the instrument base (see page 8). In case
of an instrument malfunction, turn the main power switch to the OFF position and
disconnect the instrument from the wall outlet.
Hood setup
The EVOS
™
M5000 Imaging System fits in cell culture hoods that are at least 24 in
(61 cm) deep and 36 in (92 cm) high with a 30 in (76 cm) opening. If your cell
culture hood is smaller, it may be possible to fit the instrument by turning it at a
slight angle.