102
EVOS
™
M5000 Imaging System User Guide
Configure gas
inputs
1.
Go to the
Settings
tab, then click
Incubator
to display the Incubator panel.
2.
Click
Incubator Setup
to open the Incubator Setup window.
3.
Select the appropriate options for the
Gas Inputs
that reflects your set-up for
the EVOS
™
Onstage Incubator.
•
For
Port 1
, you may select
Air
or
Premix
.
If you select
Premix
, manually enter the percentage of the
CO
2
and
O
2
to
reflect the specifics of your set-up.
•
Port 2
is reserved for Nitrogen only.
•
Port 3
is reserved for CO
2
only.
9.
Click
Apply
after you have configured the gas connections for each port.
10.
Turn on the regulators on the gas tanks. The meters on the regulators show
the tank fill on the right and gas flow on the left.
11.
Set the flow on the regulators as follows. Do not exceed 50 psi of pressure.
•
Air:
40–50 psi
•
CO
2
:
40–50 psi
•
Nitrogen:
40–50 psi