Dionex AutoTrace 280 Operator’s Manual
132
Doc. 065330-05 03/2020
2. Uncap the solvent reservoirs and fill them with the required solvents.
3. Locate the tubing weights (P/N 22163-20003), 1/4-28 PEEK nuts
(P/N 00101-07-00027), and ferrules (P/N 00101-18017) provided in
the Dionex AutoTrace 280 PFAS Ship Kit (P/N 22136-62005).
4. Insert the end of the tubing from solvent port #1 through the hole in
one of the reservoir caps. Cut the tubing length, if necessary, to avoid
slack or kinks in the solvent lines.
5. Install a 1/4-28 PEEK nut and a ferrule on one end of the tubing.
6. Secure the 1/4-28 PEEK nut at the end of the tubing to the tubing
weight.
Figure B-5. Assembly of Tubing Weight and Line
(Dionex AutoTrace 280 PFAS)
7. Cap the bottle and adjust the length of tubing until the weight touches
the bottom surface of the reservoir (see
You must provide up to five suitable containers for your solvents.
Thermo Fisher Scientific offers 1-liter glass bottles (P/N 045900) for
this purpose. The plastic bottle supplied with the Dionex AutoTrace
280 should be used for water only. Do not use the bottle for solvents;
if you do, components in the plastic bottle may leach out and
contaminate your extractions.
Tubing Weight
Tubing
1/4-28 PEEK Nut
Ferrule