5 • Service
Doc. 065330-05 03/2020
111
5. Reconnect the power cord and turn on the power.
6. Place solvent lines in the solvent reservoirs.
7. Run the Prime Solvents diagnostic method (#29).
Complete the Procedure
Run the Benchmark Test diagnostic method (#30) (see
) to verify that
the new valve works properly.
Tubing/Fitting Label
Instrument Panel Label
Valve Stator Label
V1
1 (left side)
2
V2
2 (left side)
3
V3
3 (left side)
4
V4
4 (left side)
5
V5
5 (left side)
6
V6
6 (left side)
7
W
W
8
S1
1 (right side)
14
S2
2 (right side)
13
S3
3 (right side)
12
S4
4 (right side)
11
S5
5 (right side)
10
VS
None
None (center hole)
Plug
P
1
Plug
P
9
Table 5-4. Tubing/Fitting Connections for the 12-Port Valve