5 • Service
Doc. 065330-05 03/2020
93
4. While holding the luer-lock fitting in your left hand and the syringe in your
right hand, turn the syringe counterclockwise until it disengages from the
fitting.
Mount the New Syringe
1. Insert the luer-lock fitting into the chrome end piece of the new syringe and
turn the syringe clockwise until secure.
2. Align the grooves in the chrome end piece with the tabs in the housing.
3. Gently push the syringe into place.
4. Position the syringe clamp across the chrome end piece and secure it in place
by tightening the knurled knob.
5. Gently pull the syringe plunger down until it contacts the top of the driving
arm.
6. Secure the plunger to the arm with the black thumbscrew previously removed.
Adjust the New Syringe
1. Loosen the knurled thumbscrew approximately two turns, as shown in
.