B • Installation
Doc. 065330-05 03/2020
131
8. Cap the bottle and adjust the length of tubing until the weight touches
the bottom surface of the reservoir (see
Figure B-4. Tubing Weight in Reservoir
9. Repeat
through
for each reservoir to be used.
Go on to
now to install the waste drain and tubing.
Connections to a Dionex AutoTrace 280 PFAS
To connect the solvent tubing lines on the left side of the system to the
solvent reservoirs:
1. Position the reservoirs to the left of the Dionex AutoTrace 280 PFAS,
keeping the following considerations in mind:
•
To prevent contamination, wear cleanroom gloves (lint-free,
particle-free, and oil-free) when handling tubing or making
tubing connections.
•
The reservoirs must be within reach of the tubing that extends
from the left side of the system.
•
The most precise liquid transfer occurs when the solvent
reservoirs are positioned slightly higher than the system's 12-port
valve. Thermo Fisher Scientific recommends that you do not
place the solvent reservoirs on the top of the system.