Dionex AutoTrace 280 Operator’s Manual
134
Doc. 065330-05 03/2020
2. Decide whether you want the waste tubing to extend from the left or
right side of the Dionex AutoTrace 280, and then orient the waste
drain's hose fittings to the appropriate side.
3. Install the waste drain onto the elution station's two ball studs, making
sure the waste drain snaps into place and sits flush on the base plate
(see
Figure B-8. Ball Stud and Base Plate
4. Push the waste drain toward the back of the system; it moves only
about 3 mm (1/8 in). This correctly positions the drain under the
manifold.
5. Note the 2.44-m (8-ft) length of tubing (P/N 071147) connected to the
caps of the two 12-liter waste containers (P/N 071159). One piece of
tubing is for aqueous waste; the other is for solvent waste. Screw the
caps onto the waste containers.
6. With one hand supporting the waste drain, press the free ends of the
waste tubing onto the drain's hose fittings (see
Ball Stud
Base Plate