4. Insert two new 3.15 amp IEC127 fast-blow fuses into the springs in the fuse
drawer. Press gently to fully insert the fuses into the drawer.
5. Insert the fuse drawer into the right-side panel and press until the drawer
snaps into place.
6. Reconnect the power cord and turn on the power switch.
7. Verify that the
READY
light comes on and the system operates normally.