Installation Process
Ethernet
Note that the Ethernet cable inside the analyzer can be connected to the
computer, the PLC, or the remote computer module depending on your
options and requirements. See the schedule drawings later in this manual
for wiring specifications and port locations. Additional drains and ground
wires may be connected to the extra pins in the ground blocks. Use wiring
colors and connections as follows:
Wire Color
RJ45 Pin
Number
Signal
T568A
T568B
1 T
Green/White
Orange/White
2 Transmit-
Green
Orange
3
Orange/White
Green/White
4 Unused
Blue
Blue
5 Unused
Blue/White
Blue/White
6 Receive-
Orange Green
7 Unused
Brown/White
Brown/White
8 Unused
Brown Brown
PLC
Connect the PLC I/O wires inside the enclosure to the desired signals,
noting the wire numbers used. To facilitate service and troubleshooting,
label the individual wires with the bank, terminal, and signal to which they
are connected.
The signals available depend on the modules installed in your PLC.
Consult the PLC documentation for specific wiring information for the
individual modules. Twelve banks are available for PLC modules. As
factory configured, the banks are populated as follows.
40
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific