Antaris EX-Z2 Specifications
Electrical
All electrical signals (power, alarm, and I/O) must enter and exit through
either a cable gland with armored cable or through a sealed conduit. All
wiring must comply with local codes.
Plumbing and pipe fitting
Each system requires a dedicated, switched, branch circuit electrical service,
fiber optic cabling for sample data collection, and clean, dried nitrogen or
air for purge and cooling. Depending on your application, optional
Ethernet (for remote computer and spectrometer control) and analog and
digital PLC transmission wiring also may be used.
Item
Specification
Electrical
enclosure: 0.5 –inch NPT (3) male
supply line: 0.5 –inch NPT (3) female
enclosure: 0.75 –inch NPT (2) male
supply line: 0.75 –inch NPT (2) female
cable gland with armored cable or sealed conduit
Enclosure purge
enclosure: 0.5-inch NPT male
supply line: 0.5-inch NPT female
Cooler purge
enclosure: 0.5-inch NPT female
supply line: O.5-inch NPT male
Fiber optic
spectrometer: SMA 905 threaded connection
cable: SMA 905 nut connector
1 meter tail required for spectrometer connection and
serviceability. Label with connection and sampling location.
Thermo Fisher Scientific
Antaris EX Safety, Installation, and Site Preparation Guide
91