Antaris EX-Z1 Specifications
Weight and Mounting
Weight
Model
Metric
English
Antaris EX-Z1
137 kg
300 lb
Floor stand
35 kg
75 lb
Wall mounting
Each system must be mounted to a structural member capable of
supporting four times the weight of the instrument.
All 6 bolts are required to safely secure the instrument to the wall.
The instrument must be accurately secured within 2 degrees of vertical.
The wall must be perpendicular to the floor within 2 degrees with respect
to horizontal.
For anchoring the enclosure, we recommend using high-quality steel
10 mm (3/8 in) bolts that are capable of withstanding the following forces:
Force
Specification
Shear force
4715 N (1060 lb)
Tensile force
2800 N (630 lb)
Pull-out force
2800 N (630 lbs)